Hanna Instruments HI 9828 User Manual
Page 47

47
6.4 CONDUCTIVITY
• From GLP data main menu select the “Conduc-
tivity” option and press OK. This menu allows
to view data for conductivity, actual conductiv-
ity and salinity calibrations.
• All the information about the last conductivity calibration appears: calibrated
point, cell constant value, calibration type (conductivity, actual conductivity or
salinity), time and date.
• Use the arrow keys to scroll the last 5 memorized calibrations.
• For conductivity calibration GLP data the following screens are available: con-
ductivity, actual conductivity, salinity.
NO
NO
NO
NO
NOTE
TE
TE
TE
TE
The C letter near the conductivity calibration indicates a custom point,
while the H means HANNA standard value.
If no conductivity calibration has been performed, a warning message
advise the user. Press OK to return to the previous screen.
If the selected calibration is a factory calibration, the meter shows the
message “Factory calibration”.
6.5 ATMOSPHERIC PRESSURE
• From GLP data menu select the “Atmospheric pressure” option and press OK.
• All the information about the last atmospheric pressure calibration appears:
custom calibration point, time and date.
• In case of restoring factory calibration data, the
display will show the warning message “Fac-
tory calibr. restored”.
• Use the arrow keys to scroll the last 5 memo-
rized calibrations.
NO
NO
NO
NO
NOTE
TE
TE
TE
TE
If no atmospheric pressure calibration has been performed, a warning
message advise the user. Press OK to return to the previous screen.
If the selected calibration is a factory calibration, the meter shows the
message “Factory calibration”.