Troubleshooting

FTB-500

295

Solving Common Problems

The screen does not keep its

calibration.
OR
The screen is hard to

calibrate properly.

The screen was calibrated

using the tool provided with

Windows instead of the tool

provided with ToolBox.

From the Control Panel,

go to Hardware and
Sound
> Tablet PC
Settings
.

Tap Reset.

If you need to perform

further adjustments to

the calibration, from

ToolBox, tap System
Settings
> Touch
Screen Calibration
.

The USB device I have just

connected is not working.

The device has not been

detected.

Disconnect, then reconnect

the device.
Turn off the unit, then turn it

on again.

You do not have the proper

driver for this device.

Ensure that you have the

right driver (can be provided

with the device itself).

The unit does not recognize

a test module.

Module application is not

installed.

Install the corresponding

application using the Update

Manager application.

Defective module.

If the unit recognizes other

modules, the faulty module

could be defective. Return it

to EXFO for repairs.

Module is not supported on

your unit.

Refer to the unit’s technical

specifications for the

complete list of supported

modules.

ExpressCards are not

recognized at insertion.

You might not have the

latest ExpressCard driver.

Install the most recent

ExpressCard driver available

for your card.

Problem

Possible Cause

Solution