Troubleshooting

82

FOT-930

Solving Common Problems

Unable to change

wavelength in Power Meter

pane.

Power meter receives

auto-wavelength (Auto-ID)

signal.

Change source’s emitting

mode.

Deactivate wavelength

detection.

Unable to change dB unit

or reference power in
Power Meter pane.
OR
Unit or reference value you

changed are replaced by

other values after a while.

Power meter receives

reference power value from

compatible source.

Wait a few seconds until

power value is received, then

retry.

When measuring ORL, you

get Too Much Power.

Measured reflected power

is higher than emitted

power. The ORL reference

is incorrect.

Always perform ORL

reference before each ORL

measurement.

When measuring ORL, you

get ORL exceeds.

Reflection is below

sensitivity (Max ORL) of

ORL meter.

Perform an ORL zero

measurement to increase the

sensitivity of the detector.
Ensure mandrel is correct and

that patchcord and

connectors are in good

condition.

Problem

Possible Cause

Solution