15 Troubleshooting

Solving Common Problems

Problem

Possible Cause

Solution

The unit does not turn on.

You did not press

long enough.

AC adapter/charger not

connected.

Main batteries

discharged.

Weather too cold.

Press for 2 seconds.

Connect the AC

adapter/charger.

Charge batteries by

connecting the AC

adapter/charger.

The display is almost blank

at power-up.

Brightness may need some

adjustment.

Press

to adjust brightness

properly.

The date returns to
1900-01-01.

Clock battery is dead.

Replace the clock battery

(see Replacing Batteries on

page 77).

Batteries do not charge as

expected.

Temperature is too

high.

Battery is incorrectly

connected.

Ensure temperature is

within specifications.

Ensure battery is

connected properly.

A battery status LED is

yellow.

Battery is defective.

Contact EXFO or replace the

battery.

During offset nulling, you

get the following message:
“Light detected during

nulling.”

Light reaches at least one

detector (power meter or
F

AS

T

ES

T).

Ensure protective caps are

tightly screwed on F

AS

T

ES

T

and power meter ports and

perform the nulling again.
Do not use rubber cover.

When using the power

meter, you get Power Too
Low
or Power Too High.

Power of the signal

received at the power

meter port is outside its

measurement range.

Check the connections.

Ensure you use the proper

fiber and connector type and

that you use your power

meter within specifications.