OPERATION (CONTINUED)

A

BORTING THE

D

ECONTAMINATION

A

UTO

C

YCLE

To abort the decontamination cycle perform the following steps:

End of procedure

Press and hold both the Decontamination Cycle Switch and the Normal

Operation Switch for approximately 4 seconds.

a. The CO

2

display will show the letters “Ab” to indicate that

SCO6AD has successfully aborted the cycle.

b. The Decon Activated indicator light will extinguish, though the

Decon Cycle light will remain illuminated.


The CO

2

display will then show the letters “Cdn”.

a. This indicates that the incubator is cooling down.

b. The amber Decontamination Cycle and green Normal Operation indicators will flash

alternatingly throughout the cool down stage.

c. The door will remain locked during the cool down stage. This

is a safety feature to safeguard against injuries caused by
exposure to hot chamber surfaces and shelving.


The cool down period will end the temperature reaches 50

°C.

a. The CO

2

display will show the letters “PAS” when the abort

has concluded.


Unlock the door and return the SCO6AD to normal operations.

a. Press and hold the Operation Selector Switch.
b. The CO

2

display will return to showing the concentration level

in the chamber.

c. The door will unlock, accompanied by an audible mechanical

sound.

d. The amber Decon Cycle indicator will deactivate. The green

Normal Operation light will illuminate.