To abort the decontamination cycle perform the following steps:
End of procedure
Press and hold both the Decontamination Cycle Switch and the Normal
Operation Switch for approximately 4 seconds.
a. The CO
2
display will show the letters “Ab” to indicate that
SCO6AD has successfully aborted the cycle.
b. The Decon Activated indicator light will extinguish, though the
Decon Cycle light will remain illuminated.
The CO
2
display will then show the letters “Cdn”.
a. This indicates that the incubator is cooling down.
b. The amber Decontamination Cycle and green Normal Operation indicators will flash
alternatingly throughout the cool down stage.
c. The door will remain locked during the cool down stage. This
is a safety feature to safeguard against injuries caused by
exposure to hot chamber surfaces and shelving.
The cool down period will end the temperature reaches 50
°C.
a. The CO
2
display will show the letters “PAS” when the abort
has concluded.
Unlock the door and return the SCO6AD to normal operations.
a. Press and hold the Operation Selector Switch.
b. The CO
2
display will return to showing the concentration level
in the chamber.
c. The door will unlock, accompanied by an audible mechanical
sound.
d. The amber Decon Cycle indicator will deactivate. The green
Normal Operation light will illuminate.