OPERATION (CONTINUED)
M
UTING THE
A
UDIBLE
T
EMPERATURE
A
LARM
An audible and visual high or low deviation alarm will activate if the incubation chamber
temperature deviates by 1
°C above or below the temperature set point. The low deviation audible
alarm has a delay of fifteen minutes. This prevents the low alarm from sounding whenever the
doors are opened, causing a temporary drop in temperature.
To mute an active high or low deviation alarm, press and hold either the Up or
Down arrow on the Temperature Control panel, until the amber Mute LED
illuminates and the audible alarm shuts off.
The audible alarm component will remain muted for the duration of the current
temperature deviation. The visual alarm indicator will remain illuminated.
Any new deviation of
±1°C or greater will reactivate the audible alarm.