beautypg.com

Connect the co, Supply to the incubator, Installation (continued) – Shellab SCO6AD User Manual

Page 14: Onnect the, Upply to the, Ncubator

background image

14 |

P a g e

INSTALLATION (CONTINUED)

C

ONNECT THE

CO

2

S

UPPLY TO THE

I

NCUBATOR

Note: Always use medical grade CO

2

. Use of non-medical grade CO

2

risks introducing

contaminants into the chamber, may damage the incubator,

and will void the incubator’s

warranty. Failure to use the in-line HEPA filter also voids the warranty.

Always install the in-line HEPA filter (Part Number 2800525) provided with the SCO6AD tubing
kit between the gas source and the incubator. The filter is directional. Make sure that it is installed
correctly for the flow direction (see page 41). Failure to install the filter will decrease the lifespan of
the SCO6AD, and risks allowing contaminants into the incubation chamber.

During normal operations the incubator uses only small quantities of CO

2

. Precise regulation of the

gas input flow is vital for the incubator performance. Always use a two-stage CO

2

pressure

regulator. Some single-stage regulators have 2 gauges. Make certain your regulator is a two-
stage regulator.

Connect the CO

2

regulator to a medical grade CO

2

cylinder or source.

Set the CO

2

cylinder regulator or wall source regulator to 15 - 20 Pounds per Square

Inch (psi). Do not exceed 20 psi.

PSI

Megapascals

Kilopascals

Bar

15 - 20 psi

0.103 - 0.137 Mpa

103.42 - 137.89 Kpa

1.03 - 1.378 bar

Please see the

Pressure Units Conversion table

on page 43 in the Unit Specifications section for

the formulas for converting psi into other units of pressure measurement.

Attach the black 1/8 OD tubing of the tubing kit to the regulator or wall source.

Attached the clear 5/16 OD to the CO

2

In Port on the back of the incubator.

Do not open the regulator the CO

2

incubator at this time.