beautypg.com

Calibrate the co, Display, E 30 – Shellab SCO6AD User Manual

Page 30: Operation (continued), Alibrate the, Isplay

background image

30 |

P a g e

OPERATION (CONTINUED)

C

ALIBRATE THE

CO

2

D

ISPLAY

Note: The incubator must be stabilized at its calibrated operational temperature and CO

2

set points

before the CO

2

display can be accurately calibrated. The incubator must supplied with a feed

of 15

– 20 psi of CO

2

throughout this procedure.

Always use a CO

2

reference sensor regularly calibrated to a concentration accuracy of 0.1%. For

best results use a digital gas analyzer with a sample collection tube.

The incubator should be calibrated at the concentration level it will be set to during your incubation
application.

Wait until the unit has been stabilized at its CO

2

set point for one hour with no changes in

concentration on the CO

2

display panel before starting this procedure. Failure to wait until the unit is

stabilized will prevent an accurate calibration from being carried out.

Procedure continued on next page

Connect a digital CO

2

analyzer sample tube to the sample port, located on

the OTL control panel.

Measure the current CO

2

percentage (%) inside the chamber with the

reference sensor.

a. Compare the reference reading with the percentage (%)

shown on the incubator CO

2

display panel.

b. If the readings are the same, or fall within the acceptable

range of your laboratory protocol the incubator is now
calibrated for CO

2

. The Calibrate CO

2

Display procedure is

complete.


If there is a difference between the two (2) readings, and that offset falls

outside the acceptable range of your laboratory protocol, place the CO

2

display into calibration mode. See the next step.