Page 15: Installation (continued)

background image

15 |

P a g e

INSTALLATION (CONTINUED)

I

NSTALLATION

C

LEANING AND

D

ISINFECTION

If required by your laboratory protocol, clean and manually disinfect the incubator chamber and
shelving components. Cleaning and disinfecting during installation reduces the risk of
contamination. The chamber and shelving were cleaned at the factory, and a decontamination
cycle was run to verify the cycle functionality. However, Sheldon Manufacturing cannot guarantee
that the incubator was not exposed to contaminants during shipping.

Please see the

Cleaning and Disinfecting

entry on page 39 in the User Maintenance section for

information on how to clean and disinfect without damaging the incubator or its components. A
decontamination cycle will be run as part of the Incubator setup in the Operation section.

D

EIONIZED AND

D

ISTILLED

W

ATER

Do not use deionized water to clean or humidify the incubator. Use of deionized water may
corrode metal surfaces and voids the warranty. Sheldon Manufacturing recommends the use of
distilled water in the resistance range of 50K Ohm/cm to 1M Ohm/cm, or a conductivity range of
20.0 uS/cm to 1.0 uS/cm, for cleaning and humidifying applications.

S

HELVING

I

NSTALLATION

Remove all protective wrappings from shelving components prior to installing:

Install the shelf standards.

a. Align the keyhole slot of the standard with the mounting

peg on the side of the chamber wall.

b. Mount the shelf standard.

Install the shelf slides.

a. Insert the shelf slide into the shelf standard using a

rocking motion.

b. The shelf slide will sit level when correctly

installed.

Install the shelves.

a. Slide into position with the bent ends facing

up in the back and down in the front.

Install the humidity pan.

a. Place the copper anti-microbial token in the humidity pan.

b. Secure the token using the clip on the bottom of the pan.

c. Place the pan on the chamber floor.