OPERATION (CONTINUED)
Temperature Calibration Continued
If the temperature readings of the incubator and the reference device still fall outside your
laboratory protocol after three calibration attempts, contact
for assistance.
Note: At least three attempts may be required to sucessfully calibrate a unit that is more than 2
°C
out of calibration.
End of Procedure
S
ET THE
O
VER
T
EMPERATURE
L
IMIT
The incubator must be operating at the user-selected set point temperature, and must be stable for
at least one hour prior to setting the OTL. Perform the following steps to set up the Over
Temperature Limit system for use:
End of Procedure
Example
If you have not done so already, turn the Set Over Temperature Limit
control dial clockwise to the maximum position.
Turn the Over Temperature Limit control dial counterclockwise until the
red Over Temp Limit Activated light illuminates.
Slowly turn the dial clockwise until the Over Temperature Limit
Activated light turns off. Stop turning the control.
Leave the OTL dial set slightly above the activation point. The Over
Temperature Limit is now set approximately 1
˚C above the current
chamber air temperature.