OPERATION (CONTINUED)

Preparation Procedure Continued

Complete the following procedures in the Operation section:

a.

Set the Temperature Set Point

page 24

b.

Calibrate the Temperature Display

page 25

c.

Set the Over Temperature Limit

page 27

d.

Set the CO

2

Set Point

page 29

e.

Calibrate the CO

2

Display

page 30

f.

Run a decontamination cycle

page 33

End of Preparation Procedure

H

UMIDIFYING THE

I

NCUBATOR

Filling the humidity tray with distilled water and placing it on the heated incubator floor will raise the
humidity level of the chamber to approximately 90

– 95% relative humidity.

Always use the antimicrobial copper token when humidifying the incubator to slow the
growth of microbiological populations in the tray.

Regularly clean and disinfect tray.

Refill as needed, and change the water in the tray at least once per week.

Use of chemical disinfectants added to the tray may alter the surface tension of the water.
This may significantly reduce the rate of evaporation and impact the humidity level of the
incubator chamber.

Never use deionized water.

Ensure that the unit is humidified when performing temperature adjustments and
calibrations.