7.1.3 Auto Zero
The automatic zero point adapter (auto zero) provides continuous
correction for minor weight variations such as contamination on the
weighing platform.
In the event that small quantities are removed or added to the material to
be weighed, incorrect weighing results can be displayed due to the
“stability compensation“. (e.g. slow flow of liquids from a container placed
on the balance, evaporating processes).
When apportioning involves small variations of weight, it is advisable to
switch off this function.
Available settings:
- Auto zero disabled
- Auto zero enabled
7.1.4 LO threshold
Entering threshold for automatic operation
The next measurement will not be saved until the weighing display falls
below the net value set for the LO threshold.
Tap
input window will appear.
Enter value and take over by
.
7.1.5 Last digit display accuracy
Lower display accuracy results in faster display.
Tap
< Last digit >
The menu will appear.
Always
All digits on
Never
Last digit off
When stable Last digit on for
stable weighing value