2 application specific settings – KERN & SOHN KET-TM User Manual
Page 116

116
KEN-TM / KET_TM-BA-e-1420
12.2 Application specific settings
Tap icon
in the bottom left corner of the window or call it up from the menu
(For details see chap.8.2).
The application specific settings will be displayed.
Tap button.
Function deactivated
Function activated
Explication:
Ask for multiplier
Enable query, by how many multiples the amount of the recipe
has to be created (such as double the amount = factor 2). The
target weights of the components to be weighed will then be
adjusted accordingly.
Ask for numer of cycles
Query about how often the recipe has to be processed.
Confirm batching
ingredients manually
Each individual component must be imported manually by
pressing the
PRINT key. Tare every time you add a component
by pressing the
TARE key.
Automatic tare
Automatic import of individual components after stability has
been established.
Automatic taring on start of recipe and every time a component
is added.
Ingredient control
When “
Check Component“ is enabled the editing window
component.
The editing window is used to enter the component’s code with
the help of a bar scanner.
Portion weighing
To reach the target weight, enable “Dosing of Components in
Random Dosages“ mode.
Record printout
Automatic log printout after recipe completion