3 detailed information about interface logs, 1 zeroing, 2 taring – KERN & SOHN KET-TM User Manual
Page 168

168
KEN-TM / KET_TM-BA-e-1420
16.3 Detailed information about interface logs
16.3.1 Zeroing
Befehl:
Z CR LF
Possible responses:
Z
˽
A CR LF
Z
˽
D CR LF
Command accepted; executing command
Execution of previously started command completed
Z
˽
A CR LF
Z
˽
^ CR LF
Command accepted; executing command
Command valid but zero setting range was exceeded
Z
˽
A CR LF
Z
˽
E CR LF
Command accepted; executing command
Time limit for stabilisation of weighing scale display was exceeded
Z
˽
CR LF
Command valid but can currently not be executed
16.3.2 Taring
Command:
T CR LF
Possible responses:
T
˽
A CR LF
T
˽
D CR LF
Command accepted; executing command
Execution of previously started command completed
T
˽
A CR LF
T
˽
^ CR LF
Command accepted; executing command
Command valid but tare range was exceeded
T
˽
A CR LF
T
˽
E CR LF
Command accepted; executing command
Time limit for stabilisation of weighing scale display was exceeded
T
˽
I CR LF
Command valid but can currently not be executed