4 configure display unit – KERN & SOHN KET-TM User Manual
Page 196

196
KEN-TM / KET_TM-BA-e-1420
21.4 Configure display unit
Before configuring a display unit make sure you know the required configuration data,
see chap. 21.2. The weighing system is equipped with a service switch protected
service level for entering configuration data.
Call up service level:
Switch on device with
To call up the service level you have to operate the service switch.
Use a suitable instrument for pushing the service switch
(such as thin screwdriver, Ø approx. 3 mm).
Fig.: Position
service switch
Attention:
To operate the service switch of calibrated weighing systems you will first have to
destroy the seal.
After destruction of the seal the weighing system must be re-verified by an
authorised agency and a new verification wire/seal mark fitted before it can be
reused for applications subject to verification.
Call up menu with
and the complete menu (See chap. 2.3.1) incl. service level
will be displayed. The two service blocks of the service level are described below:
<
General >
<
Manufacturer parameters >