KERN & SOHN KET-TM User Manual
Page 202

202
KEN-TM / KET_TM-BA-e-1420
21.4.2 Adjusting a weighing system
Tap
< Adjustment >
The menu will be displayed
Determine initial
mass
Adjust zero point
Calibration
How to carry out
adjustments
Initial mass
A/D transformer at zero
Calibration factor Adjusting co-efficient
Adjustment
weight
For mass of required
adjustment weight see list of
parameters “Configuration”.
How to carry out adjustment:
•
Prepare the required adjustment weight. The weight to be used depends on the
capacity of the scale. Carry out adjustment as near as possible to the scale’s
maximum weight. Info about test weights can be found on the Internet at:
•
Observe stable environmental conditions. Stabilisation requires a certain warm-
up time.
•
When using several platforms follow the sequence of steps below for each
platform.