List of parameters “Configuration“ (Display example single range scale 6 kg):
Parameters
Default
setting
Adjustable
settings
Description
Readability 1
0.001
0.001, 0.002, 0.005,
0.02, 0.05, 0.1, 0.2, 0.5,
1, 2, 5, 10, 20, 50
Readability
First weighing range
Readability 2
0.001
0.001, 0.002, 0.005,
0.02, 0.05, 0.1, 0.2, 0.5,
1, 2, 5, 10, 20, 50
Readability
First weighing range
Readability 3
0.001
0.001, 0.002, 0.005,
0.02, 0.05, 0.1, 0.2, 0.5,
1, 2, 5, 10, 20, 50
Readability
First weighing range
Weighing range
max
6.009
Weighing range
[Max + 9e]
Weighing range 2
0.000
Transition temperature
First weighing range
2. Weighing range
For single range scale select setting “0“
Weighing range 3
0.000
Transition temperature
First weighing range
3. Weighing range
For single and dual range scale select
setting “0“
Adjustment weight
1.000
Weighing value
“Adjustment weight“
Adjusting unit
kg
g, kg, lb
Weighing unit “ Admustment weight“
G-cor
Set correction value for geometric value
Auto zero range
0,25
0,1 -5
Autom. Zero tracking:
Zero setting range in unit of
readability
Stability range
5 d
0,1 -5
Rest position range in unit of
readability
Stabilisation time
0 – 20
Time interval for
stability control
Correction time
Not documented
Correction range
Not documented
Start of
control mass
0
switched off
Start up zero setting range
1
-5% to + 15 %
2
±
20 %
Digit marker
no
Represent last decimal place in bracket,
YES/NO
yes
+
After entering configuration data you have to carry out an adjustment or
linearization.
For implementation see subsequent chapters 21.4.2 / 21.4.3.