18 User’s Guide

Switch #

Switch Name

Switch

Settings

Result

3

OPA_EN_H

OFF

Sensor Excitation control is

enabled continuously

4

DIO7_OUT

ON

5

SENSOR_EXC

OFF

Sensor Excitation Output

disabled; Sensor Excitation

Bypass enabled

6

EXC_BYPASS

ON

CAUTION: Switch 5 and Switch 6 should NEVER be in the “ON”

position at the same time. This could damage the NI DAQCard.

Enable current limit monitoring

The internal circuitry of the Sensor Excitation Output provides

a Current Limit Flag. The flag lets you determine if the devices

attached to the Sensor Excitation are drawing more current than the

Sensor Excitation is capable of supplying.

You can set the PFI5 Output on the NI DAQCard to monitor the

Current Limit Flag. Note that when the switches are set to do this,

the PFI5 is no longer available through the 50-pin D connector.

Switch #

Switch Name

Switch

Settings

Result

1

I_FLAG

ON

Current Limit Monitor

Enabled; PFI5 Output used

to monitor voltage current

2

PFI5_OUT

OFF

Note: For the pod to function correctly, make sure Switch 1 and Switch

2 are never in the “ON” position at the same time.

Disable current limit monitoring and reconnect the PFI5
signal to the 50-pin connector

To make the PFI5 available again through the 50-pin D connector

again, set the switches as shown in the following table.

Switch #

Switch Name

Switch settings

Result

1

I_FLAG

OFF

Current Limit Flag is

enabled; the Current

Limit Monitor is disabled

2

PFI5_OUT

ON