5.10
Trigger Special Commands– RW, TT, TS, DT, TW, TI, HT
Remark: These commands are
only available in firmware 102.183.v.1.10.
Note: All setups should be stored with the WP command before power off.
RW
Re-Trigger Window
[ 2500sub03 ]
Set the re-trigger window in counts (digits) without decimal point.
If the weight relative to the
current average value changes by more than the RW value the average cycle will be
restarted using TT as measure time.To automatically issue the re-trigger command, the time
period over which an increase of weight average is measured has to be defined by using the
command DT.
Permitted values are 0 …65535 counts.
Master (PC / PLC) sends
Device responds
Result
RW R+65535
Request: RW=65535
RW_500 OK
Setting: RW=500d
Factory default: 65535
TT
Re-Trigger Time
[2500sub04 ]
Set the re-trigger time in milliseconds [ms].
Re-trigger time is the average time used by the re-
trigger function. If set to zero the re-trigger function is disabled.
Permitted values are 0 …65535 ms.
Master (PC / PLC) sends
Device responds
Result
TT T+65535
Request: TT=65535
TT_300 OK
Setting: TT=300ms
Factory default: 65535
TS
Re-Trigger Stop
[ 2500sub08 ]
Set the re-trigger stop in counts (digits) without decimal point.
In case of a (TS) decrease in
weight relative to the current average value the re-trigger function is stopped.
Permitted values are 0 …65535 counts.
Master (PC / PLC) sends
Device responds
Result
TS T+00000
Request: TS=00000
TS_480 OK
Setting: TT=480d
Factory default: 00000
[ The Re-Trigger function can also be stopped by sending 01 instead of 80 with
RPDO2 [00 01]. ]