(3)
2PULSE
Accumulation of charges Is started and ended through the trigger pulse input to the DC IN/SYNC terminal or
trigger signal input (LVDS) to the DIGITAL terminal, the vertical sync signal is reset and field images are
output.
Trigger pulse level
(DC IN/SYNC terminal)
Trigger pulse fetch timing
■When the trigger pulse fetch
Trigger pulse width
Trigger pulse interval
■When the trigger pulse fetch
Trigger pulse width
Trigger pulse interval
: (Low Level): Less than 0.5V , (High Level): 3.4V to 5V
: Rising edge / Falling edge / Pulse width selectable
timing is set to pulse width
: More than 16.7ms(FLD/FRM setting; FLD)
More than 33.4ms(FLD/FRM setting: FRM)
: More than 16.7ms(FLD/FRM setting: FLD)
More than 33.4ms(FLD/FRM setting: FRM)
timing is set to rising edge or falling edge
: More than 2/is
: More than 16.7ms(FLD/FRM setting: FLD)
More than 33.4ms(FLD/FRM setting: FRM)
45