77

AD-1653 Density Determination Kit




Example


Temperature Water

density

0

°C

0.99984

g/cm

3

10

°C

0.99970

g/cm

3

20

°C

0.99820

g/cm

3

30

°C

0.99565

g/cm

3



AD-1683 DC Static Eliminator

The eliminator is able to dissipate static electricity on a

charged sample.

The eliminator does not use a ventilator and can be used

in precision powder weighting.



AX-USB-9P-EX USB Converter

Adds a COM port to a PC.
Enables bi-directional communication between the PC and the balance when a USB driver is

installed.

Can use serial communication software such as WinCT on a PC without COM ports.
An RS-232C cable is provided to connect the USB converter to the balance.
















Weigh in air

Weigh in water - Weigh in the air

x water density = sample density

10.0000 g

10.0000 g - 9.5334 g

x 0.9970 g/cm

3

= 21.4 g/cm

3

To USB port

on a PC

RS-232C

on the GH series

balance

USB serial

converter

RS-232C cable provided