Details of the function table – A&D Weighing GH-252 User Manual
Page 36
34
10.2.
Details of the Function Table
Class Item
and
Parameter
Description
0
Fast response, sensitive value
FAST
1
MID.
Cond
Condition
2
Slow response, stable value
SLOW
Can be changed by response
adjustment.
0
Stable range is ±1 digit
1
#1
5t-b
Stability band width
2
Stable range is ±3 digits
The stabilization indicator illuminates
when the display fluctuation is within
the range.
0
OFF
1
Normal
2
Strong
trc
Zero tracking
3
Very strong
Keeps zero display by tracking zero
drift.
0
5 times/second
5pd
Display refresh rate 1
10 times/second
Period to refresh the display
0
Point (.)
pnt
Decimal point
1
Comma (,)
Decimal point format
0
OFF
ba5fnc
Environment
Display
p-on
Auto display-ON
1
ON
Turns on the weighing mode display
when the AC adapter is connected.
Cl adj
Clock
Refer to "10.7. Clock and Calendar Function"
The time and date are added to the
output data.
0
Key mode
Accepts the PRINT key only when
the display is stable.
1
Auto print mode A
(Reference = zero)
2
Auto print mode B
(Reference = last stable value)
Outputs data when the display is
stable and conditions of ap-p, ap-b
and the reference value are met.
prt
Data output mode
3
Stream mode /
Interval memory mode
With data 0, outputs data
continuously; with data 2, uses
interval memory.
0
Plus only
Displayed value>Reference
1
Minus only
Displayed value ap-p 2 Both Regardless of displayed value 0 10 digits 1 100 digits ap-b 2 1000 digits Difference between reference value and displayed value #2 0 Not used 1 Stores weighing data dout data 2 Stores calibration data Related items: prt, int, d-no, 5-td, info : Factory settings. #1 The unit of minimum display is digit. Example: If 1 mg display is selected using the RANGE key for the GH-300, 1 mg is one digit. #2 Usable minimum display of the balance is one digit. Example: In gram display, one digit is 0.01 mg for the GH-252 and 0.1 mg for the GH-300.
Auto print polarity
Auto print difference
Data output
Data memory