Calibration test report using an internal mass – A&D Weighing GH-252 User Manual
Page 50
48
Calibration test report using an internal mass
Note Calibration test does not perform calibration.
Key operation
1 Press and hold the
CAL
key to display
CC in
and release the key.
2
CC
is displayed and the balance is tested automatically.
3 The zero point is measured and the weight value is displayed for a few seconds.
4 Internal mass is weighed and the weight value is displayed for a few seconds.
5 If GLP output is used,
glp
is displayed and the calibration test report is output.
6 The balance returns to weighing mode after this test.
Commad
This calibration test report can be performed with commad TST.
Setting of "info 1"
Setting of "info 2"
AD-8121 printer format
General format
Space, ASCII 20h
Terminator, CR , LF or CR
CR
Carriage return, ASCII 0Dh
LF
Line feed, ASCII 0Ah
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration test type
Zero point value
Target weight value
Target weight
Signature
-----------
A
-
&
-
D
MODEL
-----
GH-300
S/N
-----
01234567
ID
------
LAB-0123
DATE
--
2004/07/01
TIME
----
12:34:56
CAL.TEST(INT.)
ACTUAL
-------
0.0000
--
g
----
+200.0002
--
g
TARGET
----
+200.0000
--
g
SIGNATURE
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-