12. Data Memory
Data memory is a function to store weighing data and calibration data in memory. The data
stored in memory are available for outputting at one time to a printer or personal computer.
Excluding date and time
Up to 200 sets
Weighing data
Including date and time
Up to 100 sets
Calibration report
Internal
calibration
External calibration
Calibration test report
Internal
test
calibration
External test calibration
Last 50 sets
12.1.
Notes on Using Data Memory
To use the memory function, set the "Data memory (
data
)" parameter and "Time/Date
output (
5-td
)" parameter of the function table. Refer to "10. The Function Table" for details on
setting the data memory.
For weighing data, the data contents to be stored and the storage capacity depend on the
"Time/Date output (
5-td
)" parameter setting.
If a different type of data exists in memory,
"Clr"
blinks the upper
left of the display. For example, you want to store weighing data
but calibration data or unit mass data remains in memory.
Under such a condition, before storing data, delete the data in memory as follows:
Releasing "Clr" or "err"
1 Press and hold the
key until
SClr "n"o
with
"no"
blinking is displayed, then release
the key.
2 Press
the
RE-ZERO
key to display
SClr "g"o
with
"go"
blinking.
The type of data stored in memory appears in the upper left of the display as shown below:
Weighing data without time and date
-d-
Weighing data with time and date
d-t
Calibration report
Hi5
3 Press
the
key to delete all the data in memory.
4 The balance displays
end
and returns to the weighing mode.
Upper left of
the display