beautypg.com

Description of the item "data output mode, Key mode, Auto print modes a and b – A&D Weighing GH-252 User Manual

Page 40: Stream mode

background image

38

10.4.

Description of the Item "Data Output Mode"

The parameter setting of "Data output mode (

prt

)" applies to the performance when the

"Data memory (

data

)" parameter is set to "

2

" (to store the weighing data) and when the

data is transmitted using the RS-232C interface.

Key mode

When

the

PRINT

key is pressed with the stabilization indictor turned on, the balance

outputs or stores the weighing data and the display blinks one time.

Required

setting

dout

prt 0

Key

mode

Auto print modes A and B

When the displayed value is stable and the conditions of "Auto print polarity", "Auto print
difference" and reference value are met, the balance outputs or stores the weighing data.

When

the

PRINT

key is pressed with the stabilization indictor turned on, the balance

outputs or stores the data and the display blinks one time.

Auto print modes A

Example

For weighing each time a sample is placed and removed, with "

ar-d

"

set to "

1

" (to adjust zero after the data is output).

Required

setting

dout

prt 1

Auto print mode A (reference = zero)

dout

ap-p

Auto print polarity

dout

ap-b

Auto print difference

dout

ar-d 1

Zero after output

Auto print modes B

Example

For weighing while a sample is added.

Required

setting

dout

prt 2

Auto print mode B (reference = last stable value)

dout

ap-p

Auto print polarity

dout

ap-b

Auto print difference

Stream mode

The balance outputs the weighing data continuously regardless of the display condition.
When the display refresh rate is set to 5 times / second (5pd 0), the data output rate is
also set to the same 5 times / second. The display does not blink in this mode. The
interval memory mode is used when the "Data memory (

data

)" parameter is set to "

1

"

(to store the weighing data).

Example

For monitoring data on a computer.

Required

setting

dout

prt 3

Stream mode

dout

data 0

Data memory function is not used

ba5fnc 5pd

Display refresh rate

5if

bp5

Baud rate

Caution The balance may not transmit the data completely at the specified refresh

rate, depending on the baud rate or data added to the weighing data such
as time, date and ID number.

This manual is related to the following products: