A&D Weighing GH-252 User Manual
Page 37
35
Class Item
and
Parameter
Description
0
Every measurement
1
2 seconds
2
5 seconds
3
10 seconds
4
30 seconds
5
1 minute
6
2 minute
7
5 minute
int
Interval time
8
10 minute
Interval time in the interval memory
mode when using prt 3, data 1
0
No output
d-no
Data number output
1
Output
Refer to "12. Data Memory"
0
No output
1
Time only
2
Date only
5-td
Time/Date output
3
Time and date
Selects whether or not the time
or date is added to the weighting
data. Refer to "10.7. Clock and
Calendar Function" for details.
0
No output
5-id
ID number output
1
Output
Selects whether or not the ID
number is output.
0
No pause
pU5e
Data output pause
1
Pause (1.6 seconds)
Selects the data output interval.
0
Not used
at-f
Auto feed
1
Used
Selects whether or not automatic
feed is performed.
0
No output
1
AD-8121 format
info
GLP output
2
General data format
Selects GLP output method.
Refer to "11. ID Number And GLP
Report "
for details.
0
Not used
dout
Data output
ar-d
Zero after output
1
Used
Adjusts zero automatically after data
is output
0
600
bps
1
1200
bps
2
2400
bps
3
4800
bps
4
9600
bps
bp5
Baud rate
5
19200 bps
0
7 bits, even
1
7 bits, odd
btpr
Data bit, parity bit
2
8 bits, none
0
CR LF
Crlf
Terminator
1
CR
CR: ASCII code 0Dh
LF: ASCII code 0Ah
0
A&D standard format
1
DP format
2
KF format
3
MT format
4
NU format
5if
Serial
interface
type
Data format
5
CSV format
Refer to "10.5. Description of Item
"Data Format".
: Factory settings.
Caution
The balance may not transmit the data completely at the specified refresh rate, depending
on the baud rate or data added to the weighing data such as time, date and ID number.