8.2.

Automatic Self Calibration

Automatic self calibration due to changes in temperature

This function automatically calibrates the balance when the balance detects an ambient

temperature change. If GLP output is selected in the function table, the balance outputs the
calibration report or stores the data in memory. Automatic self calibration functions even if
the display is turned off (standby state). Refer to "9-1. Permit Or Inhibit" for the operation.

Caution

When using automatic self calibration, do not place something on the weighing pan.
If something is on the weighing pan, the balance decides that it is in use and does

not perform automatic self calibration.

When weighing a light sample or installing the balance in a system, turn off

automatic self calibration.

Note

When turning on the balance with nothing on the pan, if a sample heavier than 0.5
g is placed on the pan, the balance detects the state that a sample is placed on the
pan and does not perform the automatic self calibration.

The

mark

is " prior notice indicator of automatic self calibration".

When the balance detects a change in ambient temperature, this
indicator blinks and automatic self calibration is required. If the
balance is not used for several minutes with this indicator blinking,
the balance performs automatic self calibration. The environment will
affect the time that the indicator blinks.

The balance is measuring calibration data. Do not allow vibration or drafts

to affect the balance while this indicator is displayed. After calibration,
the balance returns to indicate the previous display.

Advise

The balance can be used while the indicator blinks. But, it is recommended that to

maintain the accuracy, stop using the balance and confirm that there is nothing on the
pan and allow the balance to perform self calibration.

8.3.

One-Touch Calibration

This function calibrates the balance using the internal mass.
The only operation required is to press the

CAL

key.

1 Connect the AC adapter and warm up the balance for at least one

hour with nothing on the weighing pan.

2 Press

the

CAL

key to display

Cal in

.

3 The balance performs calibration using the internal mass.

Do not allow vibration or drafts to affect the balance.

4 The balance displays

end

after calibration.

If the

GLP output is set,

the balance displays

glp

and outputs the calibration report using

the RS-232C interface or stores the data in memory. Refer to "GLP
output (

info

)" and "Data memory (

data

)" of the function table.

5 The balance will automatically return to the weighing mode after

calibration.

6 Confirm weighing accuracy using calibration test (

CC in

).

GLP output