Page 56
54
Automatic self calibration can not be used while the interval memory mode is active.
The following commands can not be used during data storage.
Q
Query command for weighing data.
S
Query command for stable weighing data.
SI
Query command for weighing data.
SIR
Query command for continuous weighing data.
Setting the function table
Parameter settings for each output mode are as follows:
Item
Mode
Data output
mode
Auto print polarity,
difference
Data memory
function
Interval time
Key mode
prt 0
Not used
data 1
Auto print mode A
prt 1
data 1
Auto print mode B
prt 2
ap-a 0-2
ap-b 0-2
data 1
Not used
Interval memory mode
prt 3
Not used
data 1
int 0-8
Additional
parameter
settings, as follows:
No
d-no 0
No
5-td 0
Data number
Yes
d-no 1
Time only
5-td 1
No
5-id 0
Date only
5-td 2
ID number
Yes
5-id 1
Time and date
Both
5-td 3
Enabling the data memory function
1 Press and hold the
RANGE
key until
ba5fnc
is displayed, then release the key.
2 Press
the
RANGE
key several times to display
dout
.
3 Press
the
key.
4 Press
the
RANGE
key three times to display
data 0
.
5 Press
the
RE-ZERO
key to display
data 1
.
6 Press
the
key to store the setting.
7 Press
the
CAL
key to return to the weighing mode.