17. Specifications

GH-120

GH-200

GH-300

GH-202

GH-252

220 g

250 g

Weighing capacity

120 g

220 g

320 g

51 g

101 g

220.0084 g 250.0084 g

Maximum display

120.0084 g

220.0084 g

320.0084 g

51.00009 g 101.00009 g

0.1 mg

Minimum display

0.1 mg

0.01 mg

0.1 mg

Repeatability
(Standard deviation)

0.1 mg

0.2mg

0.02 mg

0.03 mg

±0.2 mg

Linearity

±0.2 mg

±0.3 mg

±0.03 mg

±0.10 mg

Approx. 3.5 seconds

Stabilization time

(Typical at

FAST

)

Approx. 3.5 seconds

Approx. 8 seconds

Sensitivity drift,

10°C to 30°C / 50°F to 86°F

±2 ppm/°C

Operating environment

5°C to 40°C, 85%RH or less (No condensation)

Display refresh rate

5 times/second or 10 times/second

Minimum unit

mass

0.1 mg

Countin

g

mode Number of

samples

10, 25, 50 or 100 pieces

Minimum 100%

reference mass

10.0 mg

Percent

mode Minimum 100%

display

0.01 %, 0.1 %, 1 % (Depends on the reference mass stored.)

Interface

RS-232C with Windows Communication Tools Software WinCT

External calibration
weight


100g

50g

200g
100g


300g
200g
100g


200g
100g

50g
20g

Weighing pan

φ90 mm

External dimensions

217(W) x 442(D) x 316(H) mm

Power supply &
AC adapter type

Power consumption: Approx. 11VA (supplied to the AC adapter )

Confirm that the adapter type is correct for the local voltage and

power receptacle type.

Mass Approx.

8.2

kg

The operating environment does not include excessive change of ambient temperature,

humidity, vibration, drafts, magnetic fields and static electricity.

The internal mass may change due to corrosion or other damage caused by the operating

environment, or due to aging. Check the internal mass periodically and correct the internal
mass value if necessary.