beautypg.com

Function switch and initialization, Permit or inhibit – A&D Weighing GH-252 User Manual

Page 31

background image

29

9. Function Switch and Initialization

9.1.

Permit or Inhibit

The balance stores parameters that must not be changed unintentionally (Example:

Calibration data for accurate weighing, Data for adapting to the operating environment,
Control data for the RS-232C interface). There are five switches for the purpose of protecting
parameters. Each switch can select either "permit" or "inhibit". "Inhibit" protects parameters
against unintentional operations.


1 Press

the

ON:OFF

key to turn off the display.


2 While pressing and holding the

PRINT

key and the

RANGE

key,

press the

ON:OFF

key to display

p5

.


3 Press

the

PRINT

key. Then the balance displays the function switches.


4 Specify the switches using the following keys.

RANGE

key ...........The key to select blinking digit.

RE-ZERO

key.........The key to change the parameter for the selected switch.

0 To inhibit changes. (Can not be used.)

1 To permit changes. (Can be used.)

PRINT

key .............The key to store the new parameter and return to the weighing mode.

CAL

key.................The key to cancel current operation and return to the weighing mode.

















Function table

0 To inhibit changes to the function table.

1 To permit changes to the function table.

Calibration using the internal mass (One-touch calibration)

0 To inhibit calibration using the internal mass.

1 To permit calibration using the internal mass.

Calibration using the external weight

0 To inhibit calibration using the external weight.

1 To permit calibration using the external weight.

Automatic self calibration (due to changes of ambient temperature)

0 To inhibit automatic self calibration.

1 To permit automatic self calibration.

Internal mass correction

0 To inhibit correction.

1 To permit correction.

This manual is related to the following products: