Glp report, Calibration report using the internal mass – A&D Weighing GH-252 User Manual
Page 49
47
11.2.
GLP Report
Set the following parameters to output the report.
Refer to "15.2. Connection to peripheral equipment" for connection to an AD-8121B.
To print the report, set the "GLP output (
info
)" parameter to "
1
" and use MODE 3 of the
AD-8121B. If MODE1 is used, select temporary dump print mode by pressing the
STAT.
key of the AD-8121B.
To output the report to a personal computer using the RS-232C interface, set the "GLP
output (
info
)" parameter to "
2
".
If the time and date are not correct set the correct time and date in "Clock (
Cl adj
)" of the
function table.
Calibration report using the internal mass
Key operation
1 Press
the
CAL
key to display
Cal in
and calibrate the balance automatically.
2 If GLP output is used,
glp
is displayed and the calibration report is output.
3 The balance returns to weighing mode after this calibration.
Setting of "info 1"
Setting of "info 2"
AD-8121 printer format
General format
Space, ASCII 20h
Terminator, CR , LF or CR
CR
Carriage return, ASCII 0Dh
LF
Line feed, ASCII 0Ah
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Signature
===========
A
=
&
=
D
MODEL
=====
GH-300
S/N
=====
01234567
ID
======
LAB-0123
DATE
==
2004/07/01
TIME
====
12:34:56
CALIBRATED(INT.)
SIGNATURE
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-