Response adjustment, Automatic response adjustment – A&D Weighing GH-252 User Manual
Page 23
21
7. Response
Adjustment
This function stabilizes the weight value, reducing the influence on weighing that is caused
by drafts and/or vibration at the place where the balance is installed. This function adjusts by
automatically analyzing the environment or by hand-operation. The function has three stages
as follows :
Indicator Parameter Response
Stability
FAST
Cond 0
MID.
Cond 1
SLOW
Cond 2
Fast response, Sensitive value
Slow response, Stable value
7.1.
Automatic Response Adjustment
This function automatically updates the response adjustment by analyzing the influence of
the environment using the internal mass.
1 Press and hold the
MODE
key until
RESPONSE
is
displayed, then release the key.
2 The balance automatically sets the response characteristic.
Caution Do not allow vibration or drafts to affect the
balance during adjustment.
3 After automatic adjustment, the balance displays
1
end
1
,
returns to the weighing mode and displays the updated
response indicator for about thirty seconds.
Note
If the automatic response adjustment fails, the balance displays
CH ng
. Check the ambient
conditions such as breeze and vibration, also check the weighing pan. Then, perform the
adjustment again. To return to the weighing mode, press the
CAL
key.
If there is matter on the weighing pan, the balance displays
CH 0
. Remove them from the
pan. To return to the weighing mode, press the
CAL
key.
Advise
If the automatic response adjustment is not helpful, try "7.2. Manual Response Adjustment".
Response
indicator
Press and hold
Release
Result