beautypg.com

Response adjustment, Automatic response adjustment – A&D Weighing GH-252 User Manual

Page 23

background image

21

7. Response

Adjustment

This function stabilizes the weight value, reducing the influence on weighing that is caused
by drafts and/or vibration at the place where the balance is installed. This function adjusts by
automatically analyzing the environment or by hand-operation. The function has three stages
as follows :

Indicator Parameter Response

Stability

FAST

Cond 0

MID.

Cond 1

SLOW

Cond 2

Fast response, Sensitive value

Slow response, Stable value


7.1.

Automatic Response Adjustment

This function automatically updates the response adjustment by analyzing the influence of
the environment using the internal mass.



1 Press and hold the

MODE

key until

RESPONSE

is

displayed, then release the key.




2 The balance automatically sets the response characteristic.

Caution Do not allow vibration or drafts to affect the

balance during adjustment.


3 After automatic adjustment, the balance displays

1

end

1

,

returns to the weighing mode and displays the updated
response indicator for about thirty seconds.


Note

If the automatic response adjustment fails, the balance displays

CH ng

. Check the ambient

conditions such as breeze and vibration, also check the weighing pan. Then, perform the
adjustment again. To return to the weighing mode, press the

CAL

key.

If there is matter on the weighing pan, the balance displays

CH 0

. Remove them from the

pan. To return to the weighing mode, press the

CAL

key.


Advise

If the automatic response adjustment is not helpful, try "7.2. Manual Response Adjustment".

Response

indicator

Press and hold

Release

Result

This manual is related to the following products: