12

4-1

About Maintenance

Maintenance must be performed periodically to ensure that the static elimination performance of the
SJ-M Series is maximized. Maintenance personnel should thoroughly read the descriptions under
"Safety Information for SJ-M Series" (

page 1), and perform maintenance according to the

following points.

About Maintenance

When the SJ-M Series is used for a long period of time, the electrode probe may become dirty due to
the adhesion of dust and dirt.
If the ion level alarm indicator lights, clean the electrode probe. If the electrode probe is used in a dirty
or dusty state, static elimination performance can no longer be fully achieved, resulting in accidents or
malfunction. We recommend periodically cleaning the electrode probe (as a guideline, once every two
weeks in a regular operating environment, though this depends on the installation conditions).
If cleaning the electrode probe does not improve the static elimination performance, or the ion level
alarm indicator frequently lights, replace the electrode probe.

• The SJ-M Series uses high voltage. Before starting maintenance, be sure

to turn the power OFF. Failure to do so might result in electric shock or
malfunction.

• Do not directly touch the electrode

probe. Take care not to touch these
probes even if the power is turned OFF.
Directly touching these probes may
cause personal injury.

CAUTION

4-2

Performing Maintenance on the Electrode probe

When the SJ-M Series is used for a long period of time, the electrode probe becomes dirty due to the
adhesion of dust and dirt.
If the electrode probe is used in a dirty or dusty state, static elimination performance can no longer be
fully achieved, resulting in accidents or malfunction. Be sure to periodically perform maintenance on
the electrode probe.

Performing Maintenance on the Electrode probe

How to clean the electrode probe

Clean the electrode probe gently with a cotton swab
moistened with alcohol.

How to replace the electrode probe

Note: If cleaning the electrode probe does not improve static elimination performance, or the ion level
alarm indicator frequently lights, a probable cause is that the electrode probe has reached the end of
its service life.
Electrode replacement units for SJ-M021 (OP-51607) and for SJ-M021G (OP-75351) are available as
options.

How to remove and install the electrode probe

The electrode probe can be removed by holding the tip of
the electrode probe and turning counterclockwise, and
installed by turning clockwise.

Before removing the Electrode Unit, turn the SJ-M Series OFF.

Do not directly touch the electrode probe with your hands. Doing so may
cause personal injury. Pay attention to this when performing maintenance
on the electrode probe.

CAUTION

When replacing the electrode, install the electrode holding the thicker resin
section of the Electrode Replacement Kit.

Important

Important

5-1

Timing Charts

This section provides timing charts for SJ-M Series.

Ion generation control timing chart

Indicator states when static elimination is OFF

Normal static elimination OFF
input (terminal)>
The center LED of the ion monitor blinks (red).

Input response timing chart

Normally ON Normally OFF Normally ON Forced OFF Normally ON Normally OFF Normally ON

Static elimination
stop input (terminal)

ON

OFF

Static elimination
stop input (controller)

ON

OFF

Ion emission state

Unit indicator

Emitting

No emissions

Alarm output

ON

OFF

Forced OFF

Forced OFF

Normally ON

Normally OFF

Max. 15 ms

Max. 1 s

Static elimination
stop input (unit)

ON

OFF

High voltage output

Emitting

No emissions

Normal static elimination OFF static elimination stop input (controller)>
The three center LEDs of the ion monitor
blink (red).

5-1

Timing Charts

Ion level alarm output timing chart

Condition alarm output timing chart

Alarm output timing chart

Power

ON

OFF

Static elimination
stop input (terminal)

ON

OFF

Static elimination
stop input (controller)

ON

OFF

Alarm indicator

Blinking

OFF

Alarm output (N.C.)

ON

OFF

Ion emission

Emitting

No emissions

Ion level indicator

Max. 10s

Blinking

OFF

Ion level alarm
output (N.O.)

ON

OFF

When the ion level alarm is output,
removing the cause of the alarm can
restore the normal state. One way of
restoring the normal state is to
perform maintenance on the electrode
probe.

For details on electrode probe
maintenance, see "Performing
Maintenance on the Electrode probe"
(

page 12).

Power

ON

OFF

Alarm indicator

Blinking

OFF

Alarm output (N.C.)

ON

OFF

Ion emission

Emitting

No emissions

Condition indicator

Blinking

OFF

Condition alarm
output (N.O.)

ON

OFF

Max. 10s

Static elimination
stop input (terminal)

ON

OFF

Static elimination
stop input (controller)

ON

OFF

When the condition alarm is output,
removing the cause of the alarm can
restore the normal state. One way of
restoring the normal state is to
improve the installation environment.

Power

ON

OFF

Static elimination
stop input

ON

OFF

Alarm indicator

Blinking

OFF

Alarm output (N.C.)

ON

OFF

Ion emission

Emitting

No emissions

Max. 10 s

When the alarm is output, the normal
state can be restored by performing
one of the two available restore
methods depending on the cause of
alarm output.

For details on how to restore the
normal state, see “During an alarm
(levels 1, 2)” (

page 15).