Saving records

When the measurement stops, the record is automatically saved.

The current value when measurement is stopped (electric charge/humidity/

temperature) and the ion balance measurement time or the static elimination

time are saved.

Records are saved from number 001 to number 100. Note that if a

measurement is performed with a record number in which data has already

been saved, the existing data will be overwritten. If you press the F1 key to

cancel a measurement that has started, the unit will return to the state that it

was in prior to the start of measurement. Switch to the record number that you

will use, and then start the measurement.

Hold down the F2 key when measurement is stopped to delete the data of the

displayed record number.

z

List of screens in charge plate monitor mode

You can use the

W

and

X

arrow keys to switch between the screens.

Reference

You can only switch between record number screens when the unit is

not performing a measurement. You can switch the record number

regardless of the display mode. If the record number is 100 and you

continue to switch it forward, the record number will return to 1.

z

How to execute zero point adjustment in charge plate monitor mode

1

Select a screen on which the F3
"ZERO" key is displayed.

Switch the screen using the W/X keys, and
then check that "ZERO" is displayed for the
F3 key.

2

Execute zero point adjustment.

Direct the unit toward the target that
is grounded, and press the F3
“ZERO” key.

CAUTION

You cannot switch to the screen marked with "*" during
measurement.

No.004

9.4s

10

15s

+40

−2v

−40

0V

5

0

15: 53

MENU

START

SCALE

Ion Balance

No.012

1.1s

2

3s

0kV

−0.1

−32V

−1.0

1

0

15: 23

MENU

START

Decay

No.013

0.8s

2

3s

1.0

0.1

+

9V

0kV

1

0

15: 22

MENU

START

Decay +

No.005

V

°C

%

+

2

sec

1.5

25.3

54

9/05 9: 21

9: 21

MENU

START

ZERO

ZERO

ZERO

ZERO

Ion Balance

No.002

V

°C

%

+

32

sec

1.0

26.8

50

8/23 16: 12

15: 52

MENU

START

Decay +

No.003

V

°C

%

−24

sec

0.2

29.5

46

8/28 15: 23

15: 23

MENU

START

ZERO

Decay

No.002

V

°C

%

+

2

sec

1.5

25.3

54

9/05 9: 21

9: 21

MENU

START

ZERO

Ion Balance

(1) IB value display

W

/

X

keys*

(2) IB graph display

(3) Decay+ value

display

W

/

X

keys

(6) Decay- graph

display

(5) Decay- value

display

(4) Decay+ graph

display

z

Record number switching

You can use the

S

and

T

keys to switch the

record number.

W

/

X

keys

W

/

X

keys*

W

/

X

keys

W

/

X

keys*

No.005

V

°C

%

+

2

sec

1.5

25.3

54

9/05 9: 21

9: 21

MENU

START

ZERO

Ion Balance

3.

Specifications

Specifications

*1 When measured with a response time of 0.8s, this is ±100 V or less in high-precision mode

and ±1 kV or less in wide-range mode, and ±10% (indicated value) in other ranges.

*2 The measurement accuracy range is ± 30 kV.
*3 At 25°C and with a humidity of 50%RH
*4 SK-H055 is necessary.
*5 When ±100 V or less. Other ranges are ±10% (indicated value).
*6 The laser classification is implemented on the basis of IEC 60825-1 following the

requirements of Laser Notice No. 50 of the FDA (CDRH).

*7 10 to 60%RH when using the SK-H055.

Dimensions

„

SK-H050

SK-H050

Charging

potential

measurement

Measurement mode

High-precision mode

Wide-range mode

Measurement distance

25

mm

100

mm

Measurement range

±2 kV

±50 kV

Measurement accuracy

*1

±10 V

±100 V

*2

Minimum display unit

0 to 999 V : 1

V

1.00 to 9.99 kV : 0.01

kV

10.0

kV or more : 0.1 kV

Sampling rate

Approx. 1.4 ms

Temperature

measurement

Measurement range

0 to 40°C

Measurement accuracy

*3

±1°C

Minimum display unit

0.1°C

Sampling rate

1

s

Humidity

measurement

Measurement range

10% to 85% RH

Measurement accuracy

*3

±5%

RH

Minimum display unit

1%

RH

Sampling rate

1

s

Charge

plate

monitor

function

*4

Ion

balance

measure-

ment

mode

Ion balance

measurement range

±1 kV

Measurement

accuracy

*5

±10 V

Measured voltage

minimum display unit

1 V

Static

elimina-

tion time

measure-

ment

mode

Charge voltage

±1400 V

Static elimination time

minimum display unit

0.1

s

Measurement time

0 to 99 s

Laser class

Class 1 Laser Product

(IEC60825-1: 2007

FDA(CDRH)Part1040.10

*6

)

PC interface

USB 2.0 Full Speed

Ratings

Power supply voltage

AA alkaline dry cell battery × 2

Operation time

8 hours (in charge potential

measurement mode)

Environmental resistance

Ambient temperature

0 to 40°C (no freezing and no

condensation)

Ambient humidity

*7

10 to 85%RH (no condensation)

Materials

SK-H050: PC-ABS, PC, SUS

SK-H055: PC, SUS, PTFE, PVC

Weight

SK-H050: Approx. 240 g
SK-H055: Approx. 220 g

77.6

183.3

62.8

34.5

47.7

31.6

Unit: mm