5

E SK-H

„

Initialization

1

From the measurement screen,
press the "MENU" (F1) key to
display the MENU screen.

2

Select "Initialize" and then press
the "ENTER" (F2) key.

3

On the "Initialize" menu, the
"Results" and "Factory default"
items are displayed. Select the
item that you want to initialize,
and then press the"ENTER" (F2)
key.

Results: Only the measured data will be
initialized.
Factory default: The measured data
and all settings other than the date and
time will be reset to their default values.

Mode

Response
Utility
Initialize

Static Meter

0.8s

>
>

15: 19

BACK

ENTER

MENU

Mode
Response
Utility

Initialize

Static Meter

0.8s

>
>

15: 19

BACK

ENTER

MENU

2.

Operation Menus and

Measurement Procedures

Setup Screen

„

Switching to the setup screen

From the measurement screen, press the "MENU" (F1) key. The MENU screen is
displayed.

„

List of setting items

„

Setting details

z

Mode

Set the measurement mode.

z

Response

Set the measurement response time.

* This can only be set during charging potential measurement mode.

z

Utility

Configure settings such as the date and time and the operation sound - for a
variety of functions.

* Area Scaling Function
This function corrects errors in the measured value due to size differences of the
measurement targets by setting a correction value according to the size of the
target.

Setting item

Details

Mode

Select the measurement mode.

Response

Set the response time.

Utility

Configure functional settings such as the date and time, operation sound,

and backlight.

Initialize

Initialize the measured data and settings.

Setting item

Details

Default

value

Static Meter

Charge potential measurement mode: The electric charge of the

target will be measured.

{

Charge Plate

Charge plate monitor mode

With an IMU (ionizer monitoring unit) connected, the static

elimination speed and ion balance of the static eliminator will be

measured.

* To use this mode, an optional ionizer monitoring unit is required.

Setting item

Details

Default

value

Response

0.1s/0.2s/0.4s/0.8s/1.6s/3.2s/6.4s

0.8s

Setting item

Details

Default

value

Date

Set the date and time.

Sound

Turn the operation sound on or off.

Setting item: ON/OFF

ON

Backlight

Set the display backlight. When this is set to "Auto," the

backlight will remain on for a fixed length of time after an

operation is performed.

Setting item: Auto/ON/OFF

Auto

Sleep

When no operations are performed for the set time, the unit

will automatically turn off.

Setting item: OFF/3min/5min/10min/15min/30min

OFF

Laser Pointer

Turn the laser emission on or off.

Setting item: ON/OFF

ON

Area Scaling

Set the area scaling.
Setting Items: Near/FAR (setting range: 0.1 to 1.5)

Near: 1.0
Far: 1.0

Mode

Response
Utility
Initialize

Static Meter

0.8s

>
>

15: 19

BACK

ENTER

MENU