Initial reset (reset to the default state), Switching the display unit, Setting mode display – KEYENCE GT-70A Series User Manual
Page 7: Function setting mode
7
GT-M-E
■
Initial reset (Reset to the default state)
This function cancels all the functions you set and allows the
equipment to be reset to the default state.
• The initial reset can be set on the main display only.
• The initial reset cannot cancel "span adjustment results".
1
When in the main display, while
pressing the [MODE] button, press
the [SET] button five times.
"rSt.no" will appear.
2
Press the top/bottom Arrow buttons
to select "rSt.YES".
3
Press the [MODE] button.
"rSt.End" will appear.
The initial reset is completed.
■
Switching the display unit
The display unit can be switched from the initial reset screen.
This section explains how to switch from "mm" display to "inch" display.
1
Display "rSt.YES" from the initial reset screen.
Refer to "Initial reset" for more details on how to display "rSt.YES".
2
Pressing the [MODE] button, press the [SET] button for 2
seconds or more.
3
Use the up and down Arrow button to
select "inch".
4
Press the [MODE] button.
"rSt.End" will appear.
Switching the display unit is completed.
Setting Mode Display
■
Function setting mode
You can set the following functions in the function setting mode.
No.
Item
Description
Detection mode Select the detection mode.
Response
time
Longer response time makes the average
data time longer, which stabilizes the value.
Timing type
For the timing input, select "External timing"
or "Self timing (internal timing)".
Self timing
level
When the self timing is selected in
, set
the self timing level.
Self timing
delay type
When the value rises above (or falls below) the
self timing level with "Std" selected in
,
select "Fix the criterion value when the setting
time is elapsed" or "Fix the criterion value after
the display value is stabilized (static hold)".
User specified
delay time
Set the delay time when "Fix the criterion value
when the setting time is elapsed" is selected in
.
Static hold
delay stability
criterion
Set the reference of stability criterion when "Fix
the criterion value after the display value is
stabilized (static hold)" is selected in
.
Static hold
delay stability
width
Enter the optional numerals when USEr (user
specification) is selected in
.
Calculation
mode
Set the calculation method when an
expansion unit(s) is added. *The calculation
mode will appear only when an expansion
unit(s) is added.
Note
MODE
Standard
Detection mode (press and hold the [MODE] button in the main screen)
NG hold
Peak hold
Bottom hold
Peak to peak
Display for entering the setting value
Response time
MODE
External timing input
Timing type
Rising self
Falling self
MODE
Available to set only when any of Std, P-H and b-H is selected in [ Detection mode].
MODE
Display for entering the setting value
Self timing level
Available to set only when item other than t-in is selected in [ Timing type].
MODE
Static hold
Self timing delay type
Available to set when the following conditions are provided.
•
Std is selected in [ Detection mode].
•
Item other than t-in is selected in [ Timing type].
Delay timer
MODE
Display for entering the setting value
User specified delay time
Available to set only when t.d is selected in [ Self timing delay type].
MODE
Default
Static hold delay stability criterion
Available to set only when Stb.d is selected in [ Self timing delay type].
User
Display for entering the setting value
Static hold delay stability width
Available to set only when USEr is selected in [ Static hold delay stability criterion].
MODE