Hardy HI 2151/30WC Single-Scale Controller User Manual
Page 60
![background image](/manuals/657640/60/background.png)
HI 2151 Series Weight Controller
Profibus Interface Option
54
11
3
57
39
Acceptable number of
averages between 1 and 200
12
4
58
3A
Span weight value, during
calibration, must be positive
13
5
59
3B
Scale capacity value must be
positive (from 1-999,999)
14
6
60
3C
Midpoint linearity value must be
positive
15
7
61
3D
Rate-of-change time units must
be 0, 1 or 2
16
8
62
3E
Rate-of-change time base out
of range
17
10
1
65
41
Analog output not installed
18
2
66
42
Analog output request must be
between 0-4
19
3
67
43
Invalid serial port (RS232)
format request
20
4
68
44
Serial configuration values 0 o1
21
5
69
45
Baud rate request out of range
22
6
70
46
Parity request out of range
(must be 0, 1 or 2)
23
7
71
47
Stop bits must be 0 or 1
24
8
72
48
Word length must be 0 or 1
NOTE:
Description in the GSD file was shortened because of 32 character limit.
Unit Diag
Bit #
Diagnostic
Byte
Diagnostic
Bit
Decimal
#
Hex #
Description
25
11
1
73
49
Control (Hardware or Software)
must be 0 or 1
26
2
74
4A
Device Address must be
between 0-99
27
3
75
4B
Echo request must be 0 or 1
28
4
76
4C
Tare greater than scale
capacity
29
5
77
4D
Blind unit option only
30
6
78
4E
Auto zero tolerance must be
between .0001-999,999
31
7
Reserved for future use