Hardy HI 2151/30WC Single-Scale Controller User Manual
Page 32
![background image](/manuals/657640/32/background.png)
HI 2151 Series Weight Controller
Profibus Interface Option
26
Block Read Data Number 35: Instrument Identification
Byte Definitions
Byte Pos.
#
Words
Start Word # Bytes Start Byte
Block Read Number 35 (Hex 23)
Firmware revisions level (one character text, i.e.
“A”)
Firmware version (ASCII text number)
Zero Calibration analog to digital converter raw
counts
Span Calibration analog to digital converter raw
counts
Total Words/Bytes
LSB
MSB
1
2
2
2
7
0
1
3
5
1
1
4
4
4
14
0
1
2
6
10
Block Read Data Number 36: Floating Point Data
Byte Definitions
Byte Pos.
#
Words
Start Word # Bytes Start Byte
Block Read Number 36 (Hex 24)
Reserved for Future Use
Gross Weight (Floating Point)
ROC Data (Floating Point)
Total Words/Bytes
LSB
MSB
1
2
2
5
0
1
3
1
1
4
4
10
0
1
2
6
Block Read Data Number 37: Integrated Technician Status
Byte Definitions
Byte Pos.
#
Words
Start Word # Bytes Start Byte
Block Read Number 37 (Hex 25)
Excitation Monitor ON/OFF 1 = ON 0 = OFF
Excitation Monitor Error (1 - Error, 0 = OK)
Reserved for Future Use
Total Words/Bytes
LSB
MSB
LSB
MSB
1
1
2
0
1
1
1
1
1
4
0
1
2
3