Error codes, List of response/error codes – Hardy HI 2151/30WC Single-Scale Controller User Manual
Page 59
![background image](/manuals/657640/59/background.png)
Section 8: Troubleshooting
53
8.4
ERROR CODES
See Section 4.4 for information on error code setup
List of Response/Error Codes
Unit
Diag Bit
#
Diagnostic
Byte
Diagnostic
Bit
Decimal # Hex
#
Description
N/A
1
0-7
N/A
N/A
Station Status 1*
N/A
2
0-7
N/A
N/A
Station Status 2*
N/A
3
0-7
N/A
N/A
Station Status 3*
N/A
4
0-7
N/A
N/A
Diagnostic server Address
N/A
5-6
0-7
N/A
N/A
Identification Number (=2151
hex)
N/A
7
0-7
N/A
N/A
Extended Diagnostics Length
N/A
8-13
0-7
N/A
N/A
Extended Diagnostics Data
(See Below)
NOTE:
See Profibus Standard
Unit
Diag Bit
#
Diagnostic
Byte
Diagnostic
Bit
Decimal # Hex
#
Description
1
8
1
06
06
Acknowledge good data
received
2
2
21
15
Negative Acknowledge
(NACK)-illegal command
3
3
49
31
Scale is in motion (e.g. unable
to calibrate in motion)
4
4
50
32
Current weight sensed over
scale capacity (command 70)
5
5
51
33
Weight not with zero tolerance,
unable to zero
6
6
52
34
Insufficient change in weight to
calibrate span (error #18)
7
7
53
35
Decimal point places must be
between 0 and 4
8
8
54
36
Not a valid graduation size
9
9
1
55
37
Motion value must be greater
than graduation size
10
2
56
38
Zero tolerance value must be
between 0001 -999,999