Service menu options – Cannon Instrument CMRV-5000 User Manual
Page 60

54
CANNON
®
Mini-Rotary Viscometer
CMRV-5000 Instruction & Operation Manual
Version 1.0b—August, 2011; CANNON
®
Instrument Company
2139 High Tech Road • State College, PA • 16803 • USA
Service menu options
To access the service options, click Service from the VISCPRO
®
primary display and select the desired instrument.
MRV Monitor
The MRV Monitor window permits the user to view current operational
values for the CMRV instrument. These settings may be helpful to
CANNON
®
technical personnel when troubleshooting difficulties with
the CMRV.
Cleaning Service
Select Cleaning Service to open the Cleaning Service window. Then
Click the Start Cleaning button to initiate an instrument heating cycle
(50°C). The current instrument temperature and instrument status will be
displayed in the window. Click the Stop Cleaning Service button when
cleaning is completed.
Temperature Probe Calib.
The Temperature Probe Calibration window provides a software
interface for calibrating the CMRV block temperature probe. Consult
Chapter 7 of the CMRV Instruction & Operation Manual for details
regarding the software-driven probe calibration procedure.