Preparing for cmrv testing/calibration, Cleaning cycle – Cannon Instrument CMRV-5000 User Manual
Page 33
27
CANNON
®
Mini-Rotary Viscometer
CMRV-5000 Instruction & Operation Manual
Version 1.0b—August, 2011; CANNON
®
Instrument Company
2139 High Tech Road • State College, PA • 16803 • USA
CHAPTER
6
PREPARING FOR CMRV
TESTING/CALIBRATION
The procedures in this chapter should be followed when preparing for
CMRV-5000 testing or calibration.
NOTE
In the event that the protocols of your ASTM test method differ from any
of the procedures outlined in this manual, the ASTM method should take
precedence.
Providing power
To prepare for operation of the CMRV-5000 unit, turn on the POWER
switch on the front panel of the Controller. The green power indicator on
the switch will light and the left panel lights will flash once.
Checking thermometers
The ice point of the CMRV-5000 thermometers should be checked
periodically at 76 mm immersion, which corresponds to the depth of
immersion in the CMRV-5000 aluminum block.
Small deviations in the ice point (less than 0.4°C) should be noted and
added to (or subtracted from) calibration readings at all other tempera-
tures (See section on temperature calibration). If corrections larger than
0.4°C are needed, there may be a problem with the thermometer. Exam-
ine the thermometer for a bubble of gas in the lower reservoir, a bubble
of mercury in the upper reservoir, or a break in the mercury column. For
information on joining separated mercury columns see APPENDIX C.
Cleaning cycle
Clean the rotors and all viscometric cells as described in the following
procedure.
CAUTION
When handling the rotors, be careful not to damage the rotor tips. Dam-
age to the tips will cause erroneous test results and may damage the
cells.
CAUTION
When operating the CMRV, make sure that the rotor and cell numbers
coincide (use the #1 rotor in cell #1, the #2 rotor in cell #2, and so on).
Failure to do so may diminish test accuracy and void the cell calibration
constants.
Cleaning procedure
1. Remove the Plexiglas
®
cover from the CMRV-5000 and place it away
from the cleaning solvents.