Appendix g - communications – Brookfield DV-I Prime User Manual
Page 64
Brookfield Engineering Laboratories, Inc.
Page 64
Manual No. M/07-022-D0613
Appendix G - Communications
DV-I PRIME
Serial and Analog Outputs
-
2
3
4
5
6
7
8
9
1
No Connection
Transmit Data
(TxD)
No Connection
Serial Ground
Analog Ground
Analog % Torque
(Note 1)
Analog Temperature
(Note 2)
Cable Sense
Serial Ground
Notes:
1. This is a 0-1 volt d.c. output where 0 volts corresponds to 0% torque and 1
volt corresponds to 100 % torque with a resolution of 1 millivolt (0.1%).
2. This is a 4.00 volt d.c. output where 0 volts corresponds to -100°C and 4.00
volts corresponds to +300°C with a resolution of 1 millivolt (0.1°C).
Figure G-1
Analog Output:
The analog outputs for % torque is accessed from the 9-pin connector located on the rear panel
of the DV-I PRIME. The pin connections are shown in Figure G-1.
The output cable (Part No. DVP-96B) connections are:
Red
Wire: Temperature Output (with temp. option)
Black
Wire: Temperature Ground (with temp. option)
White
Wire: % Torque Output
Green
Wire: % Torque Ground
NOTE:
Please contact Brookfield Engineering Laboratories or your local dealer/distributor
for purchase of the DVP-96B analog output cable.
When using the Brookfield Printer Cable (Brookfield Part No. DVP-81), the output rate is 1.0
times per second. The DV-I PRIME uses the following RS-232 parameters to output these strings:
Baud Rate
9600
Data Bits
8
Stop Bits
1
Parity
None
Handshake
None