beautypg.com

Appendix g - communications – Brookfield DV-I Prime User Manual

Page 64

background image

Brookfield Engineering Laboratories, Inc.

Page 64

Manual No. M/07-022-D0613

Appendix G - Communications

DV-I PRIME

Serial and Analog Outputs

-

2

3

4

5

6

7

8

9

1

No Connection

Transmit Data

(TxD)

No Connection

Serial Ground

Analog Ground

Analog % Torque

(Note 1)

Analog Temperature

(Note 2)

Cable Sense

Serial Ground

Notes:
1. This is a 0-1 volt d.c. output where 0 volts corresponds to 0% torque and 1

volt corresponds to 100 % torque with a resolution of 1 millivolt (0.1%).

2. This is a 4.00 volt d.c. output where 0 volts corresponds to -100°C and 4.00

volts corresponds to +300°C with a resolution of 1 millivolt (0.1°C).

Figure G-1

Analog Output:

The analog outputs for % torque is accessed from the 9-pin connector located on the rear panel

of the DV-I PRIME. The pin connections are shown in Figure G-1.

The output cable (Part No. DVP-96B) connections are:

Red

Wire: Temperature Output (with temp. option)

Black

Wire: Temperature Ground (with temp. option)

White

Wire: % Torque Output

Green

Wire: % Torque Ground

NOTE:

Please contact Brookfield Engineering Laboratories or your local dealer/distributor

for purchase of the DVP-96B analog output cable.

When using the Brookfield Printer Cable (Brookfield Part No. DVP-81), the output rate is 1.0

times per second. The DV-I PRIME uses the following RS-232 parameters to output these strings:

Baud Rate

9600

Data Bits

8

Stop Bits

1

Parity

None

Handshake

None