Brookfield DV-I Prime User Manual
Page 18
Brookfield Engineering Laboratories, Inc.
Page 18
Manual No. M/07-022-D0613
If the
ARROW
key is pressed just once and then released, the characters
RPM
will blink for three
seconds, then will cease blinking resulting in no change to the speed entry.
NOTE:
The speed selection process remembers the last value of scrolled-to speed so that
the next time you initiate a speed change (by pressing an
ARROW
key), the DV-I
PRIME will begin its scroll display from the last entered value.
The last scrolled-to-speed does not necessarily have to be the same as the speed at which the
DV-I PRIME is currently running. The user may operate at a given speed and pre-set the DV-I
PRIME to the next desired speed before that speed will be used. For example, if the DV-I PRIME
is currently running at 10 RPM and was previously scrolled to 20 RPM, a single press of either
ARROW
key would result in the Figure II-10 screen display:
cP 872.0
S01
10 RPM20
% 87.2
Scrolled to Selected Speed
Operating Speed
Figure II-10
Pressing the
SET SPEED
key would cause the DV-I PRIME to begin running at 20 RPM.
If motor is off, pressing the
MOTOR ON/OFF
key immediately starts the DV-I PRIME running at
the last scrolled-to-speed
. If you had been running at 10 RPM, pressed
MOTOR ON/OFF
and then
re-started the DV-I PRIME by pressing
MOTOR ON/OFF
once again, you would again be running
at 10 RPM. However, if while the motor was off you had scrolled to a new speed of 20 RPM,
pressing the
MOTOR ON/OFF
key would start the DV-I PRIME running at 20 RPM.
DV-I PRIME remembers last speed selected when power is shut down.
NOTE:
During both spindle or speed selection and scrolling operations, the DV-I PRIME
will continue to calculate and display viscosity (cP) and torque (%).
II.4 Autorange
The
AUTO RANGE
key allows you to determine the maximum calculated viscosity (full scale reading)
possible with the current spindle/speed setting. Pressing the key at any time will cause the current
viscosity display to change and show that maximum viscosity. The screen torque display will
now display a flashing “%100.0” to indicate this special condition. This maximum viscosity and
flashing %100.0 value will be displayed for as long as the
AUTO RANGE
key is depressed. Figure
II
-11 shows the
AUTO RANGE
function for the situation where the No. 3 RV spindle is rotating at
20 RPM. The Full Scale Range is 5000 cP (or 5000 mPa
.
s).
cP 5000
S03
20 RPM
%100.0
Figure II-11