beautypg.com

Brookfield DV-I Prime User Manual

Page 46

background image

Brookfield Engineering Laboratories, Inc.

Page 46

Manual No. M/07-022-D0613

A.5 Verifying Calibration

1. Determine the appropriate sample volume. Refer to Table A-1 to determine the correct sample

volume required for the spindle to be utilized.

2. Select a Brookfield Viscosity Standard fluid that will give viscosity readings between 10% and

100% of Full Scale Range. Refer to Appendix B for viscosity ranges of cone spindles; ranges

listed apply to CPE cones.

Do not use a silicone viscosity standard fluid with a viscosity value greater than 5000 cP with a

Cone/Plate. Brookfield offers a complete range of mineral oil viscosity standards suitable for

use with Cone/Plates for viscosities above 5,000 cP or shear rates above 500 sec

-1

; see Table E-1

and E-2 in Appendix E for a list of available fluids.

It is best to use a viscosity standard fluid that will be close to the maximum viscosity for a given

cone spindle/speed combination.

Example:

LVDV-I PRIME Viscometer, Cone Spindle CPE-42, Brookfield Silicone

Viscosity Standard having a viscosity of 9.7 cP at 25°C.

At 60 RPM, the Full Scale Viscosity Range is 10.0 cP. Thus, the viscometer reading should be

97% torque and 9.7 cP viscosity ± 0.197 (closer to ±0.2) cP. The accuracy is a combination of

viscometer and fluid tolerance (refer to Interpretation of Calibration Test Results at the end

of Appendix E).

3. With the motor off, remove the sample cup and place the viscosity standard fluid into the cup.

Cone Part No.

Sample Volume

CPE-40

0.5 mL

CPE-41

2.0 mL

CPE-42

1.0 mL

CPE-51

0.5 mL

CPE-52

0.5 mL

Table A-1

4. Attach the sample cup to the viscometer and allow sufficient time for the sample, cup and cone

to reach temperature equilibrium.

5. Turn the motor on. Set the desired speed(s). Measure the viscosity and record the reading in

both % torque and centipoise (cP).

NOTE:

The cone spindle must rotate at least five (5) times before a viscosity reading is

taken.

6. Verify that the viscosity reading is within the allowable 1% deviation, as explained earlier, for

the specific viscosity standard fluid(s) that you are using.

* The CPE designation on the cone spindle indicates use with Electronic Gap Setting Cone/Plate

Viscometers/Rheometers only.