Brookfield DV-I Prime User Manual
Page 10
Brookfield Engineering Laboratories, Inc.
Page 10
Manual No. M/07-022-D0613
I.4 Specifications
Speeds:
0.0, 0.3, 0.6, 1.5, 3, 6, 12, 30, 60,
(rpm)
0.5, 1, 2, 2.5, 4, 5, 10, 20, 50, 100
Weight: Gross Weight 20 lb
9 kg
Net Weight 17 lb
7.7 kg
Carton Volume 1.65 cu ft 0.05 m
3
Operating Environment:
0°C to 40°C Temperature Range (32°F to 104°F)
20% - 80% R.H.: non-condensing atmosphere
Analog Torque Output:
0 - 1 Volt DC (0 - 100% Torque)
Viscosity Accuracy:
±1% Full Scale Range in Use (See Appendix D for range
calculation)
Viscosity Repeatability:
0.2% of Full Scale Range in Use
Tp
Temperature Sensing Range: -100°C to +300°C (-148°F to +572°F)
Tp
Temperature Accuracy:
±1°C: -100°C to +149°C
±2°C: +150°C to +300°C
Tp
Analog Temperature Output: 0-4 Volt DC (10 mV/°C)
Electrical Certifications:
Conforms to CE Standards:
BSEN 61326:
Electrical equipment for measurement, control and laboratory use - EMC
requirements
BSEN 61010-1:
Safety requirements for electrical equipment, for measurement, control
and laboratory use
Notice to customers:
This symbol indicates that this product is to be recycled at an appropriate collection center.
Users within the European Union:
Please contact your dealer or the local authorities in charge of waste management on how to
dispose of this product properly. All Brookfield offices and our network of representatives and
dealers can be found on our website: www.brookfieldengineering.com
Users outside of the European Union:
Please dispose of this product according to your local laws.