Brookfield DV-I Prime User Manual
Page 54
Brookfield Engineering Laboratories, Inc.
Page 54
Manual No. M/07-022-D0613
Table D-2
lists the model codes and spring torque constants for each viscometer model.
VISCOMETER MODEL
TORQUE
CONSTANT TK
MODEL CODE ON DV-I
PRIME SCREEN
LVDV-I PRIME
0.09373
LV
2.5xLVDV-I PRIME
0.2343
2.5LV
5xLVDV-I PRIME
0.4686
5LV
1/4 RVDV-I PRIME
0.25
1/4RV
1/2 RVDV-I PRIME
0.5
1/2RV
RVDV-I PRIME
1
RV
HADV-I PRIME
2
HA
2xHADV-I PRIME
4
2HA
2.5xHADV-I PRIME
5
2.5HA
HBDV-I PRIME
8
HB
2xHBDV-I PRIME
16
2HB
2.5xHBDV-I PRIME
20
2.5HB
Table D-2
The full scale viscosity range for any DV-I Prime model and spindle may be calculated using the
equation:
Full Scale Viscosity Range [cP] = TK * SMC * 10,000
RPM
where:
TK = DV-I Prime Torque Constant from Table D-2
SMC = Spindle Multiplier Constant from Table D-1
The Shear Rate calculation is:
Shear Rate (
1
/
Sec
) = SRC * RPM
where:
SRC = Shear Rate Constant from Table D-1