beautypg.com

Brookfield DV-I Prime User Manual

Page 54

background image

Brookfield Engineering Laboratories, Inc.

Page 54

Manual No. M/07-022-D0613

Table D-2

lists the model codes and spring torque constants for each viscometer model.

VISCOMETER MODEL

TORQUE

CONSTANT TK

MODEL CODE ON DV-I

PRIME SCREEN

LVDV-I PRIME

0.09373

LV

2.5xLVDV-I PRIME

0.2343

2.5LV

5xLVDV-I PRIME

0.4686

5LV

1/4 RVDV-I PRIME

0.25

1/4RV

1/2 RVDV-I PRIME

0.5

1/2RV

RVDV-I PRIME

1

RV

HADV-I PRIME

2

HA

2xHADV-I PRIME

4

2HA

2.5xHADV-I PRIME

5

2.5HA

HBDV-I PRIME

8

HB

2xHBDV-I PRIME

16

2HB

2.5xHBDV-I PRIME

20

2.5HB

Table D-2

The full scale viscosity range for any DV-I Prime model and spindle may be calculated using the

equation:

Full Scale Viscosity Range [cP] = TK * SMC * 10,000

RPM

where:

TK = DV-I Prime Torque Constant from Table D-2

SMC = Spindle Multiplier Constant from Table D-1

The Shear Rate calculation is:

Shear Rate (

1

/

Sec

) = SRC * RPM

where:

SRC = Shear Rate Constant from Table D-1