beautypg.com

Brookfield DV-I Prime User Manual

Page 45

background image

Brookfield Engineering Laboratories, Inc.

Page 45

Manual No. M/07-022-D0613

A.4 Making Measurements with Cone/Plate Geometry

Viscosity measurements are made on the DV-I PRIME C/P viscometer in the same way as the

DV-I PRIME viscometer with several exceptions.

1. Prepare the viscometer as is described in Section III.2.

2. Brookfield recommends that you always make cone/plate measurements with temperature

control. Be sure that the sample cup is connected to a circulating waterbath and that it

is at the target temperature. If the set temperature is far from ambient, you may wish to

preheat the spindle as well as the cup. The spindle can be preheated by removing it from

the viscometer and resting it in the cup (be careful not to scratch the measurement surfaces

on the spindle or cup). Attach the spindle to the viscometer prior to the next step.

3. Set the Gap (see Appendix A; Section A.3) Brookfield recommends that the gap be set at

the same temperature at which the measurement will be made.

4. Remove the sample cup.

5. Measure the precise volume of sample required for the spindle. See Table A-1. Sample

volume can have a great affect on the measurement. In general it is better to over fill the

gap slightly than it is to under fill. It is also beneficial to establish a method of sampling

that is repeatable to contribute to the repeatability of your viscosity measurement.

6. Insert the sample into the center of the sample cup. Avoid air bubbles when possible.

7. Attach the sample cup to the viscometer (be careful not to tilt the sample cup as this would

reposition the sample from the center).

8. Wait for thermal equilibrium. Brookfield recommends a minimum of 1 minute for thermal

equilibrium. You may want to increase this time if; there is a large difference in temperature

between the sample and the control point, or if you have not preheated the spindle.

9. Operate the viscometer (see Section II.8, Steps 3 – 6). Refer to Section III.3 for assistance

in selecting a spindle or speed.

10. Remove the sample cup at the conclusion of the test. Be careful to lower the cup so that

no contact is made between the cup and spindle.

11. Remove the spindles (Refer to Appendix A; Section A.2, Step 4).

Cleaning:

Refer to Section I.8 for general cleaning recommendations. Clean the cup and spindle

at the conclusion of each test. Be careful not to let the sample dry or harden onto the spindle or

cup since mechanical scraping may damage the measurement surfaces. Brookfield recommends

that you remove the spindle prior to cleaning.

Take care not to scratch the measurement surface of the cup or spindle during the cleaning
process. Take care not to drop the spindle; any dent on the outer edge of the spindle will
affect the measurement.

The sample cup may be provided with an embedded temperature probe. The temperature probe

connection should not be exposed to the cleaning solution or the test sample. Do not immerse

the sample cup in a cleaning solution.