Manual No. M/07-022-D0613

II.3 Speed Selection & Setting


Table II-1 shows the available speed selections.

DV-I PRIME SPEEDS SETS

Beginning

0.0
0.3

When scrolling

0.6

“UP”

1.5
3.0
6.0

12

30
60

0.0
0.5
1.0
2.0
2.5
4.0
5.0

10
20
50
100

Table II-1

NOTE

: DV-I PRIME speeds are organized to conform to the historical speed sets available on

the Brookfield Dial Reading viscometer. Speeds from 0.3-60 RPM are traditionally

found on the LVT viscometer. Speeds from 0.5-100 RPM are traditionally found

on RVT, HAT, and HBT viscometers.

To select

a viscometer speed first press either the

UP

or

DOWN ARROW

keys which will cause the

area to the right of RPM (on the bottom line) to display the currently selected speed. Figure II-9

shows the DV-I PRIME had been operating at 10 RPM, and the current selected speed is 10 RPM.

cP 872.0

S01

10RPM10

% 87.2

“Selectable” Speed

Operating Speed