Brookfield DV-I Prime User Manual

Page 25

background image

Brookfield Engineering Laboratories, Inc.

Page 25

Manual No. M/07-022-D0613

NOTE:

For the Time to Torque method, the DV-I PRIME Viscometer will retain the

last entered torque in MEMORY for use when the user elects to perform

a time to torque test again.

Tp

II.9.4 Temperature Offset

When the optional temperature probe is available with the DV-I PRIME, it is sometimes useful

to be able to adjust the temperature readout to agree with an external temperature device. This

can be accomplished utilizing the Temp Offset mode.

1. Enter the Timed Options menu by pressing the TIMED OPTION key. Use the up or

down arrow key to select Temp Offset. The following screen will be presented.

 TEMP OFFSET
 TIMED STOP

Figure II-29

2. Press ENTER. Figure II-30 will appear.

SET TEMP OFFSET?
NO

$# THEN ENTER

Figure II-30

3. Press the UP/DOWN keys for YES or NO.

4. If NO is entered, temperature offset is disabled and display will return to default screen.

5. If YES is entered, the screen in Figure II-31 appears.

ENTER TEMP
$# +0.0 OFFSET

Figure II-31

6. Use UP/DOWN arrows to scroll from +5.0 to -5.0 degrees C offset in 0.1º increments.

(When temperature units are set to Fahrenheit, offset range is -9.0 to +9.0 degrees F)

7. Press ENTER to choose Temp Offset. The offset will be confirmed by the following

screen, which will be displayed for 5 seconds, then returns to default screen.

TEMP OFFSET ENABLED

Figure II-32