Brookfield DV-I Prime User Manual
Page 25
Brookfield Engineering Laboratories, Inc.
Page 25
Manual No. M/07-022-D0613
NOTE:
For the Time to Torque method, the DV-I PRIME Viscometer will retain the
last entered torque in MEMORY for use when the user elects to perform
a time to torque test again.
Tp
II.9.4 Temperature Offset
When the optional temperature probe is available with the DV-I PRIME, it is sometimes useful
to be able to adjust the temperature readout to agree with an external temperature device. This
can be accomplished utilizing the Temp Offset mode.
1. Enter the Timed Options menu by pressing the TIMED OPTION key. Use the up or
down arrow key to select Temp Offset. The following screen will be presented.
TEMP OFFSET
TIMED STOP
Figure II-29
2. Press ENTER. Figure II-30 will appear.
SET TEMP OFFSET?
NO
$# THEN ENTER
Figure II-30
3. Press the UP/DOWN keys for YES or NO.
4. If NO is entered, temperature offset is disabled and display will return to default screen.
5. If YES is entered, the screen in Figure II-31 appears.
ENTER TEMP
$# +0.0 OFFSET
Figure II-31
6. Use UP/DOWN arrows to scroll from +5.0 to -5.0 degrees C offset in 0.1º increments.
(When temperature units are set to Fahrenheit, offset range is -9.0 to +9.0 degrees F)
7. Press ENTER to choose Temp Offset. The offset will be confirmed by the following
screen, which will be displayed for 5 seconds, then returns to default screen.
TEMP OFFSET ENABLED
Figure II-32