Brookfield Engineering Laboratories, Inc.

Page 11

Manual No. M/07-022-D0613

I.5 Setup

1. To assemble the Model S Laboratory Stand, place the upright rod into the base (refer to

assembly instructions in Appendix H). The rack gear and clamp assembly should face

the front of the base. The upright rod is held in place with a screw, which is attached from

the bottom of the base. Tighten this screw with a screwdriver.

2. Be sure that the clamp screw, VS-41Y, is loose. Insert the mounting rod on the back of

the DV-I PRIME Viscometer into the hole on the clamp assembly.

3. Tighten the VS-41Y clamp screw. Adjust the Viscometer to be as close to level as possible

while tightening the clamp screw.

4. Connect the optional RTD temperature probe to the temperature port

on the rear

panel of the DV-I PRIME, if provided.

5. The Viscometer must be level. The level is adjusted using the two Leveling Screws (VS-3)

on the base. Adjust so that the bubble level on top of the DV-I PRIME is centered within

the circle. Check level periodically during use.

6. Remove Shipping Cap
7. Make sure that the AC power switch at the rear of the DV-I

PRIME is in the OFF position.

Connect the power cord to the socket on the back panel of the instrument and plug it into

the appropriate AC line.

The AC input voltage and frequency must be within the appropriate range as
shown on the nameplate of the viscometer. The DV-I

PRIME mus

t

be earth

grounded to ensure against electronic failure!

8. For Cone/Plate models, refer to Appendix A.

I.6 Safety Symbols and Precautions

Safety Symbols

The following explains safety symbols, which may be found in this operating manual.

Indicates hazardous voltages may be present.


Refer to the manual for specific warning or caution information to avoid personal injury

or damage to the instrument.

Precautions

If this instrument is used in a manner not specified by the manufacturer, the protection

provided by the instrument may be impaired.

This instrument is not intended for use in a potentially hazardous environment.

In case of emergency, turn off the instrument and then disconnect the electrical cord

from the wall outlet.

The user should ensure that the substances placed under test do not release poisonous,

toxic or flammable gases at the temperatures to which they are subjected to during the
testing.

Tp